Triage® MeterPlus Specification Sheet

Physical
Size8.5" D x 6.25" W x 2.75" H
Weight1.5 pounds w/o batteries; 1.6 lbs with batteries
Keypad12 numeric keys, 10 function keys
Electrical6 v DC @ 1 amp
AC / DC converter or 4 AA batteries
PrinterPanasonic Thermal Printer
InterfaceRS-232 Computer Serial Port

Environmental
Temp15 C – 30 C
Humidity10% – 85%
LocationDry, flat, horizontal surface away from direct sunlight

Optical
LaserLaser Diode – 1 milliwatt
DetectorSilicone Photodiode

Memory Capacity
User ID’s600
Patient Diagnostics750
Reagent Lot Data20
QC Diagnostics200
QC Lot Data30
QC Simulator Diagnostics70

Language Options: French, English, Spanish, Italian, German

Quality Control (QC)

  • QC Internal Standard check of laser and optics performed at time of each diagnostic.
  • Meter specific QC Simulator device for testing meter alignment and calibration, with programmable mandatory frequency.
  • Optional user ID tracking. When set, requires user performing diagnostic to enter user ID prior to commencing. Checks user ID against list and expiration date.

QC lockout when:

  • QC Internal Standard check fails
  • On board control material does not fall within limits
  • External controls not performed on reagent lot being used
  • External controls not performed within programmed frequency
  • QC Simulator not performed within programmed frequency
  • User ID not valid or expired

Supervisor-only accessible screens for meter setup parameters and Diagnostic Cutoffs