Triage® MeterPlus Specification Sheet
Physical
| Size | 8.5" D x 6.25" W x 2.75" H |
| Weight | 1.5 pounds w/o batteries; 1.6 lbs with batteries |
| Keypad | 12 numeric keys, 10 function keys |
| Electrical | 6 v DC @ 1 amp AC / DC converter or 4 AA batteries |
| Printer | Panasonic Thermal Printer |
| Interface | RS-232 Computer Serial Port |
Environmental
| Temp | 15 C – 30 C |
| Humidity | 10% – 85% |
| Location | Dry, flat, horizontal surface away from direct sunlight |
Optical
| Laser | Laser Diode – 1 milliwatt |
| Detector | Silicone Photodiode |
Memory Capacity
| User ID’s | 600 |
| Patient Diagnostics | 750 |
| Reagent Lot Data | 20 |
| QC Diagnostics | 200 |
| QC Lot Data | 30 |
| QC Simulator Diagnostics | 70 |
Language Options: French, English, Spanish, Italian, German
Quality Control (QC)
- QC Internal Standard check of laser and optics performed at time of each diagnostic.
- Meter specific QC Simulator device for testing meter alignment and calibration, with programmable mandatory frequency.
- Optional user ID tracking. When set, requires user performing diagnostic to enter user ID prior to commencing. Checks user ID against list and expiration date.
QC lockout when:
- QC Internal Standard check fails
- On board control material does not fall within limits
- External controls not performed on reagent lot being used
- External controls not performed within programmed frequency
- QC Simulator not performed within programmed frequency
- User ID not valid or expired
Supervisor-only accessible screens for meter setup parameters and Diagnostic Cutoffs
|